Send E-mail to get more information: sales@pintech.com.cn
Detailed description :
AC CURRENT PROBE PT-720
1. 2000A current with high accurate measurement. Max. Clamp 52mm.
2.Isolated coaxial BNC cable, match any meter, data recorders and oscilloscopes. Can measure and show waves easily with mV signals output.3 ranges selection, can replace A621,i1000s,etc.
Current range | 0.1A~10A | 10A~200A | 200A~2000A |
Attenuation ratio | 100mV/A (10:1) | 10mV/A (100:1) | 1mV/A (100:1) |
Accuracy | ≤3%+10mV | ≤2%+5mV | ≤1%+1mV |
| Bandwidth | 10Hz~100kHz(-3dB) | ||
Max. dV/dt | 10A/μs | ||
Max. current | when >1kHz,1000A Continuous current;If ≥1200A(no longer than 20 minutes) | ||
Load impedance | ≥1MΩ and ≤47pF | ||
Rise/Fall time | ≤3.5us | ||
Working voltage | 600V rms | ||
Common mode voltage | 600V CAT III | ||
Max. Jaw opening diameter | 53mm | ||
Max. Conductor size | Cable:max. Diameter:Ф52mm;Total wiring harness;single:5х5mm,Four bus wiring harness:30х5mm | ||
Working Temperature | 0~50℃ | ||
Working Humidity | 20~75%RH | ||
| Size | 216х111х45mm | ||
Weight | 500g | ||
Manual measurement, with an error of ±2mm.